About us ] Contact us ] Literature & Tools ] ONSITE MONITORING ]                  

Indicating Glass Moisture Traps


Specialty Gases and Equipment

(514) 956-7503

Model 6420 and 6421 Series Indicating Glass Moisture Traps

Description
The Model 6420 and 6421 purifiers utilize a molecular sieve material to remove moisture from non-corrosive gases. The units are prepared under high vacuum conditions to ensure purity, and are pre-purged with high purity helium.
Models 6420 and 6421 Glass Moisture Traps

Applications
  • Removes moisture from gases used in Electron Capture, MSD, and GC/ Mass Spec applications

Design Features
  • Borosilicate glass tube housing
  • 1/8" or ¼" stainless steel Swagelok® compression tube fittings

Specifications
Maximum Operating Pressure: 100 psig
Max. Operating Temperature: 100ºC (212ºF)
Maximum Flow Rate: 2 scfm @ 100 psig
Water Removal Capacity: 16.3 grams
Efficiency: Less than 10 ppb*
Shipping Weight: 1 lb
*By actual test of product under GC flow conditions

Ordering Information
Model Description Size Fitting
6420-2S Glass Indicating Moisture Trap 70cc 1/8"SS
6420-4S Glass Indicating Moisture Trap 70cc 1/4"SS
6421-2S Glass Indicating Moisture Trap 100cc 1/8"SS
6421-4S Glass Indicating Moisture Trap 100cc 1/4"SS
6420-CL Mounting clip for 6420 and 6421
6420-R Molecular Sieve Refill for 6420 and 6421
Please order Matheson Technical Bulletin TB-318 for additional information.

 